ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20519
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Report Date
- July 10, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/09/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED WITH INVESTIGATION. REVIEW OF THE BLACK BOX REVEALED RELATED ALARMS. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE, BOLUS DELIVERIES, AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. THE FORCE SENSOR CALIBRATION WAS WITHIN SPECIFICATION. NO DAMAGE, DEFECT, OR MOISTURE WAS OBSERVED TO THE PUMP¿S INTERNAL COMPONENTS. UNRELATED TO THE COMPLAINT, A BATTERY COMPARTMENT CRACK WAS OBSERVED.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419057 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 32 YR |