FDA Adverse Event Malfunction Summary report: N

ONETOUCHPING GLUCOSEMGMTSYSTEM

MDR report key: 3942949 · Received July 17, 2014

Report

Report Number
2531779-2014-20519
Event Type
Malfunction
Date Received
July 17, 2014
Report Date
July 10, 2014
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

THE DEVICE WAS RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 09/09/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT COULD NOT BE DUPLICATED WITH INVESTIGATION. REVIEW OF THE BLACK BOX REVEALED RELATED ALARMS. THE PUMP SUCCESSFULLY COMPLETED A PRIME SEQUENCE, BOLUS DELIVERIES, AND 24-HOUR EXERCISE TEST WITHOUT ISSUE OR ALARM. THE FORCE SENSOR CALIBRATION WAS WITHIN SPECIFICATION. NO DAMAGE, DEFECT, OR MOISTURE WAS OBSERVED TO THE PUMP¿S INTERNAL COMPONENTS. UNRELATED TO THE COMPLAINT, A BATTERY COMPARTMENT CRACK WAS OBSERVED.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A PRIME (LOSS OF PRIME) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419057 ONETOUCHPING GLUCOSEMGMTSYSTEM INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 32 YR