PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 2032227-2014-04404
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 17, 2014
- Report Date
- June 17, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WI
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
INSULIN PUMP RECEIVED WITH INTERMITTENT BUTTONS DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. INSULIN PUMP RECEIVED WITH A CRACKED CASE ON DISPLAY WINDOW CORNERS. INSULIN PUMP RECEIVED WITH A MINOR SCRATCH ON LCD WINDOW. INSULIN PUMP RECEIVED WITH MISSING END CAP STICKER.
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A BUTTON ERROR ON HER INSULIN PUMP. HER BLOOD GLUCOSE WAS 88 MG/DL. THE CUSTOMER WAS ADVISED TO DISCONTINUE USE OF THE PUMP AND REVERE TO A BACK-UP PLAN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419237 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC MINIMED | MMT-723RNAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 18 YR |