FDA Adverse Event Malfunction Summary report: N

SENSOR ENLITE

MDR report key: 3942810 · Received July 17, 2014

Report

Report Number
2032227-2014-04291
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OZO
PMA / PMN Number
120010
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSPECTED THREE SEALED ENLITE SENSORS. PERFORMED BICARBONATE BUFFER TEST. ONE SENSOR FAILED PER SPECIFICATION DUE TO LOW READINGS. THE TWO REMAINING SENSOR PASSED PER SPECIFICATION WITH ACCURATE READINGS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER RECEIVED A CALIBRATION ERROR. THE BLOOD GLUCOSE READING WAS NOT GIVEN. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419200 SENSOR ENLITE OZO OZO MEDTRONIC MINIMED MMT-7008A

Patients

Seq Age Sex Outcome Treatment
1