FDA Adverse Event
Malfunction
Summary report: N
SENSOR ENLITE
MDR report key: 3942810
·
Received July 17, 2014
Report
- Report Number
- 2032227-2014-04291
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 17, 2014
- Report Date
- June 17, 2014
- Manufacturer
- MEDTRONIC MINIMED
- Product Code
- OZO
- PMA / PMN Number
- 120010
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
Additional Manufacturer Narrative · 1
INSPECTED THREE SEALED ENLITE SENSORS. PERFORMED BICARBONATE BUFFER TEST. ONE SENSOR FAILED PER SPECIFICATION DUE TO LOW READINGS. THE TWO REMAINING SENSOR PASSED PER SPECIFICATION WITH ACCURATE READINGS.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE CUSTOMER RECEIVED A CALIBRATION ERROR. THE BLOOD GLUCOSE READING WAS NOT GIVEN. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419200 | SENSOR ENLITE | OZO | OZO | MEDTRONIC MINIMED | MMT-7008A |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |