FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME INSULIN INFUSION PUMP

MDR report key: 3942725 · Received July 17, 2014

Report

Report Number
2032227-2014-03969
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

INSULIN PUMP WAS RECEIVED WITH MOTOR ERROR ALARM DURING OCCLUSION TEST AND UNABLE TO PRIME DURING PRIME TEST DUE TO FAULTY FORCE SENSOR. MOTOR PASSED MOTOR TEST. INSULIN PUMP HAD MINOR SCRATCHES ON LCD WINDOW, CRACKED DISPLAY WINDOW CORNER, CRACKED BATTERY TUBE THREADS, BROKEN BELT CLIP SLOT AT BATTERY TUBE THREADS AREA, CRACKED RESERVOIR TUBE LIP AND CRACKED RESERVOIR TUBE WINDOW.

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE INSULIN PUMP ALARMED MOTOR ERROR. THE BLOOD GLUCOSE READING IS 250 MG/DL. CUSTOMER IS UNABLE TO REWIND THE DEVICE. THE INSULIN PUMP WILL BE REPLACED. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418526 PARADIGM REAL-TIME INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-522LNAL

Patients

Seq Age Sex Outcome Treatment
1 58 YR