FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3942697 · Received July 17, 2014

Report

Report Number
2032227-2014-04229
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 15, 2014
Report Date
June 16, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

CUSTOMER REPORTED RECEIVING A MOTOR ERROR ALARM WHEN ATTEMPTING TO DELIVER A BOLUS. THE BLOOD GLUCOSE READING WAS 87 MG/DL. CUSTOMER STATED THAT THE INSULIN PUMP WAS NOT EXPOSED TO A STRONG MAGNETIC FIELD OR AN MRI. CUSTOMER STATED THAT HE WAS ABLE TO REWIND THE INSULIN PUMP. NOTHING FURTHER WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418189 PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP INSULIN INFUSION PUMP / SENSOR AUGMENTED OYC MEDTRONIC MINIMED MMT-723RNAS

Patients

Seq Age Sex Outcome Treatment
1 75 YR