FDA Adverse Event Malfunction Summary report: N

COULTER LH 500 HEMATOLOGY ANALYZER

MDR report key: 3942667 · Received July 17, 2014

Report

Report Number
1061932-2014-01679
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 19, 2014
Report Date
June 19, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K042724
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE FIELD SERVICE ENGINEER (FSE) DISCOVERED THE FLUID LEAKED FROM PINCH VALVE 37 (PV37) ON THE INSTRUMENT. THE FSE REPLACED ALL OF THE TUBING AT PV37 AND RESOLVED THE FLUID LEAK ISSUE. THE INSTRUMENT CONFORMED TO THE MANUFACTURER'S PUBLISHED PERFORMANCE SPECIFICATIONS AND WAS RETURNED TO NORMAL OPERATION. IN CONCLUSION, THE LIKELY CAUSE OF THE EVENT IS ATTRIBUTED TO THE TUBING AT PINCH VALVE PV37. A DEFINITIVE ROOT CAUSE IS UNKNOWN. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED APPROXIMATELY THREE (3) MILLILITERS OF BLUE FLUID LEAKED ON THE RIGHT SIDE OF THE INSTRUMENT WHILE PROCESSING LATRON CONTROL INVOLVING THE COULTER LH 500 HEMATOLOGY ANALYZER. THE FLUID WAS CONTAINED WITHIN THE INSTRUMENT. THE OPERATOR WAS WEARING PROTECTIVE GLOVES AND DID NOT HAVE DIRECT CONTACT WITH THE FLUID. THERE WAS NO OPERATOR INJURY OR ADVERSE EFFECT ASSOCIATED WITH THIS EVENT. NO ERRONEOUS PATIENT RESULTS WERE GENERATED AS PATIENT SAMPLES WERE NOT ANALYZED. THERE WAS NO PATIENT IMPACT ASSOCIATED WITH THIS EVENT. THE FACILITY HAS AN EXPOSURE CONTROL AND RISK MANAGEMENT PLAN IN PLACE FOR BIOHAZARD MATERIAL. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO ASSESS THE INSTRUMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418172 COULTER LH 500 HEMATOLOGY ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1