COULTER ACT DIFF ANALYZER
Report
- Report Number
- 1061932-2014-01564
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 24, 2014
- Report Date
- June 24, 2014
- Manufacturer
- BECKMAN COULTER
- Product Code
- GKZ
- PMA / PMN Number
- K973634
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ON (B)(4) 2014, THE BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND FOUND THE WBC (WHITE BLOOD CELL) BATH OVERFLOWING BECAUSE OF A SLIGHT KINK IN TUBING THROUGH PINCH VALVE LV14. THE FSE TRIMMED THE LENGTH OF THE TUBING AND THE INSTRUMENT RAN WITHOUT LEAKS. (B)(4).
THE CUSTOMER REPORTED A LEAK WHEN USING THE COULTER ACT DIFF ANALYZER. THE LEAK WAS FROM A BATH INSIDE THE INSTRUMENT. THE VOLUME OF THE LEAK WAS LESS THAN 1 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT OF LAB COAT AND GLOVES WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR CUTS. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418365 | COULTER ACT DIFF ANALYZER | COUNTER, DIFFERENTIAL CELL | GKZ | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |