FDA Adverse Event Malfunction Summary report: N

COULTER ACT DIFF ANALYZER

MDR report key: 3942640 · Received July 17, 2014

Report

Report Number
1061932-2014-01564
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
BECKMAN COULTER
Product Code
GKZ
PMA / PMN Number
K973634
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ON (B)(4) 2014, THE BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) EVALUATED THE ANALYZER AND FOUND THE WBC (WHITE BLOOD CELL) BATH OVERFLOWING BECAUSE OF A SLIGHT KINK IN TUBING THROUGH PINCH VALVE LV14. THE FSE TRIMMED THE LENGTH OF THE TUBING AND THE INSTRUMENT RAN WITHOUT LEAKS. (B)(4).

Description of Event or Problem · 1

THE CUSTOMER REPORTED A LEAK WHEN USING THE COULTER ACT DIFF ANALYZER. THE LEAK WAS FROM A BATH INSIDE THE INSTRUMENT. THE VOLUME OF THE LEAK WAS LESS THAN 1 ML AND WAS NOT CONTAINED WITHIN THE INSTRUMENT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS SENT TO THE CUSTOMER'S FACILITY TO EVALUATE THE ANALYZER. THE OPERATOR WAS WEARING PERSONAL PROTECTIVE EQUIPMENT OF LAB COAT AND GLOVES WHEN THE EVENT OCCURRED. THERE WERE NO REPORTS OF BIOHAZARD EXPOSURE TO MUCOUS MEMBRANES OR CUTS. THERE WAS NO REPORT OF ERRONEOUS TEST RESULTS ASSOCIATED WITH THIS EVENT. THERE WAS NO DEATH, INJURY OR AFFECT TO USER OR PATIENT TREATMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418365 COULTER ACT DIFF ANALYZER COUNTER, DIFFERENTIAL CELL GKZ BECKMAN COULTER NA NA

Patients

Seq Age Sex Outcome Treatment
1