FDA Adverse Event Malfunction Summary report: N

PK DISSECTING FORCEPS INSTRUMENT

MDR report key: 3942576 · Received July 17, 2014

Report

Report Number
2955842-2014-04368
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 19, 2014
Report Date
June 20, 2014
Manufacturer
INTUITIVE SURGICAL,INC.
Product Code
NAY
PMA / PMN Number
K061260
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CT
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND THE PITCH UP CABLE FRAYED AT THE DISTAL CLEVIS HUB. FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. CONDUCTOR WIRES WERE INTACT AND NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER; THE FRAYED CABLES COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT IF THE FAILURE MODE WERE TO RECUR.

Description of Event or Problem · 1

IT WAS REPORTED THAT IN CENTRAL PROCESSING, A FRAYED WIRE WAS IDENTIFIED ON THE PK DISSECTING FORCEPS INSTRUMENT. THERE WAS NO ALLEGATION OF HARM OR INJURY TO THE PATIENT AND NOTHING REPORTEDLY FELL INTO THE PATIENT.