ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT
Report
- Report Number
- 9611451-2014-00627
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 20, 2014
- Report Date
- June 27, 2014
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LTD
- Product Code
- BZE
- PMA / PMN Number
- K983112
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- OTHER
Narratives
(B)(4). METHOD: THE TWO COMPLAINT RT340 BREATHING CIRCUITS WERE RETURNED TO FISHER & PAYKEL HEALTHCARE (FPH) IN (B)(4), AND WERE VISUALLY INSPECTED. RESULTS: VISUAL INSPECTION REVEALED THAT FOR ONE DEVICE (LOT NUMBER 140324) THE EVAQUA EXPIRATORY LIMB HAD TWO HOLES ABOUT 4CM AND 27CM RESPECTIVELY AWAY FROM THE PATIENT END CONNECTOR, FOR THE SECOND DEVICE (UNKNOWN LOT NUMBER) THERE WAS A HOLE IN THE EVAQUA EXPIRATORY LIMB ABOUT 5.5CM AWAY FROM THE PATIENT END CONNECTOR. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 140324 CONCLUSION: BASED ON OUR VISUAL INSPECTION, THE SUBJECT EXPIRATORY LIMBS APPEARED TO HAVE BEEN PUNCTURED WITH A BLUNT OBJECT. ALL RT340 BREATHING CIRCUITS ARE VISUALLY INSPECTED AND PRESSURE TESTED FOR LEAKS BEFORE RELEASING FOR DISTRIBUTION. ANY BREATHING CIRCUIT WHICH FAILS IS DISCARDED. IN ADDITION, TUBE WEIGHING AND BOND STRENGTH TESTING ARE PERFORMED EVERY 15 MINUTES. IF ANY FAULTS ARE DETECTED THE WHOLE BATCH IS PLACED ON HOLD FOR INVESTIGATION. THIS SUGGESTS THAT THE SUBJECT BREATHING CIRCUIT WAS DAMAGED AFTER IT WAS RELEASED FOR DISTRIBUTION. THE KEY DIFFERENCE BETWEEN THE EVAQUA BREATHING CIRCUITS AND CONVENTIONAL BREATHING CIRCUITS IS THAT THE EXPIRATORY LIMB OF THE EVAQUA CIRCUITS SUCH AS THE RT340 IS COMPOSED OF A THIN, SEMI-PERMEABLE FILM SPECIALLY DESIGNED TO ALLOW WATER VAPOUR FROM EXPIRED VENTILATORY GAS TO PASS THROUGH. THE EVAQUA EXPIRATORY TUBE HAS A PROTECTIVE MESH WHICH PREVENTS DAMAGE TO THE WALLS OF THE TUBE, HOWEVER THE EVAQUA TUBING CAN BE MORE SUSCEPTIBLE TO DAMAGE THAN CONVENTIONAL CIRCUITS WHEN EXPOSED TO ROUGH HANDLING OR DAMAGE CAUSED BY SHARP OBJECTS AND NON-FPH CIRCUIT HANGERS. THE USER INSTRUCTIONS THAT ACCOMPANY THE RT340 ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT STATE THE FOLLOWING: PERFORM A PRESSURE AND LEAK TEST ON THE BREATHING SYSTEM AND CHECK FOR OCCLUSIONS BEFORE CONNECTING TO A PATIENT. SET APPROPRIATE VENTILATOR ALARMS. FIT ONLY THE SUPPLIED FISHER & PAYKEL HEALTHCARE CIRCUIT HANGER WITH CARE TO AVOID CIRCUIT DAMAGE.
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THEY FOUND A LEAK IN TWO RT340 ADULT BREATHING CIRCUITS DURING A VENTILATOR LEAK TEST.THIS WAS FOUND PRIOR TO PATIENT USE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418743 | ADULT DUAL HEATED EVAQUA BREATHING CIRCUIT | BZE | BZE | FISHER & PAYKEL HEALTHCARE LTD | RT340 | 140324 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |