FDA Adverse Event Malfunction Summary report: N

HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3942518 · Received July 17, 2014

Report

Report Number
1416980-2014-23105
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 24, 2014
Report Date
June 24, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4).THE DEVICE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN; THEREFORE, A DEVICE ANALYSIS COULD NOT BE COMPLETED. THIS COMPLAINT IS FOR A REPORT OF A DAMAGED PATIENT DURING THE DRAIN STAGE, WHERE THE HOME PATIENT RAN OVER THE PATIENT LINE AN CUT OFF THE PATIENT LINE. USE ERRORS AND PROPER USER INSTRUCTIONS ARE ADDRESSED IN ¿THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE¿, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE. SECTION ¿PREPARE FOR THERAPY¿ WARNS THE USER TO FOLLOW ASEPTIC TECHNIQUE TAUGHT BY YOUR DIALYSIS CENTER WHEN HANDLING LINES AND SOLUTION BAGS TO REDUCE THE POSSIBILITY OF INFECTION. SECTION ¿PREPARE FOR THERAPY¿ WARNS THE USER THAT USING DAMAGED SETS CAN RESULT IN CONTAMINATION OF THE FLUID OR FLUID PATHWAYS. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE PATIENT LINE WAS SEVERED DURING DRAIN TWO OF FOUR ON THE HOME CHOICE (HC). THE HOME PATIENT (HP) STATED THEY HAD RUN OVER THE PATIENT LINE WITH THEIR WHEELCHAIR. THE HP STATED THEY HAD CLOSED THE CLAMPS AND CAPPED OFF THE PATIENT LINE. THE HP STATED THEY WOULD START THERAPY OVER. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) ASSISTED THE HP TO END THERAPY. NO PATIENT INJURY OR MEDICAL INTERVENTION WAS INDICATED AS A RESULT OF THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418279 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 49 YR HOMECHOICE