PROGRASP FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04358
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 4, 2014
- Report Date
- June 23, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- LA
- Reporter Occupation
- OTHER
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. THE CUSTOMER REPORTED COMPLAINT WAS CONFIRMED BY FAILURE ANALYSIS. THE PITCH CABLE WAS BROKEN AT THE DISTAL CLEVIS HUB. THE CABLE SEGMENT THAT CONTAINS THE CRIMP WAS STILL INSTALLED IN THE CLEVIS. THE CLEVIS DID NOT EXHIBIT ANY DAMAGE OR WEAR MARKS. ADDITIONALLY FAILURE ANALYSIS FOUND THE INSTRUMENT DISTAL PULLEY HAD MECHANICAL INDENTATIONS/BURRS. THERE WAS AN INDENTATION AT THE EDGE OF THE DISTAL PULLEY AND VISIBLE SCRATCHES ON THE SURFACE OF THE PULLEY. EVIDENCE NOT CONCLUSIVE, BUT MECHANICAL INDENTATION/BURRS DAMAGE MAY BE DUE TO MISHANDLING/MISUSE. NO OTHER DAMAGE FOUND. THE INSTRUMENTS AND ACCESSORIES USER MANUAL SPECIFICALLY STATES: GENERAL PRECAUTIONS AND WARNINGS -HANDLE INSTRUMENTS WITH CARE. AVOID MECHANICAL SHOCK OR STRESS THAT CAN CAUSE DAMAGE TO THE INSTRUMENTS. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A MDR REPORTABLE EVENT; HOWEVER THE BROKEN PITCH CABLE WERE TO RECUR COULD CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT.
IT WAS REPORTED DURING A DA VINCI SURGICAL PROCEDURE THAT THE PROGRASP FORCEPS INSTRUMENT HAD A SNAPPED CABLE. THERE WERE NO MISSING OR FALLEN PIECES REPORTED. THERE WAS NO ALLEGATION OF ANY HARM, INJURY, OR ADVERSE OUTCOME TO A PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418265 | PROGRASP FORCEPS INSTRUMENT | ENDOSCOPIC INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420093-10 | M10131218 773 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |