FDA Adverse Event Malfunction Summary report: N

DIMENSION EXL 200

MDR report key: 3942486 · Received July 17, 2014

Report

Report Number
1226181-2014-00385
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
January 2, 2014
Report Date
July 3, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC.
Product Code
JJE
PMA / PMN Number
K073604
Removal / Correction Number
1226181-07/16/2014-003-C
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A SIEMENS CUSTOMER SERVICE ENGINEER (CSE) WAS DISPATCHED TO THE CUSTOMER SITE. THE CSE ADJUSTED THE HINGE TENSION AND DEMONSTRATED PROPER OPERATION TO THE CUSTOMER. THE CAUSE OF THE REAGENT LID FALLING ONTO THE OPERATOR WAS A MALFUNCTION OF THE REAGENT LID HINGES. THIS INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED. SIEMENS HEALTHCARE DIAGNOSTICS HAS INVESTIGATED THE INSTANCES OF DIMENSION EXL INSTRUMENT REAGENT LIDS FALLING DURING MAINTENANCE PROCEDURES. IT HAS BEEN DETERMINED THAT THE HINGE MAY LOSE ITS EFFECTIVENESS AND SLOWLY SHIFT DOWNWARD DURING MAINTENANCE PROCEDURES. CUSTOMERS IN THE UNITED STATES WERE SENT URGENT MEDICAL DEVICE CORRECTION (UMDC) 14-46 ENTITLED "DIMENSION EXL REAGENT LID HINGE ISSUE" AND CUSTOMERS OUTSIDE OF THE UNITED STATES WERE SENT URGENT FIELD SAFETY NOTICE (UFSN) 14-46 ENTITLED "DIMENSION EXL REAGENT LID HINGE ISSUE". THE UMDC/UFSN INFORM CUSTOMERS THAT THEIR CSE WILL INSPECT AND ADJUST THE TENSION ON THE REAGENT LID HINGES DURING EVERY VISIT. IF THE HINGES CAN NO LONGER BE PROPERLY ADJUSTED, THE HINGES WILL BE REPLACED. THE UMDC/UFSN ALSO INSTRUCTS CUSTOMERS TO CONTACT THE SIEMENS CUSTOMER CARE CENTER OR THEIR LOCAL SIEMENS TECHNICAL SUPPORT REPRESENTATIVE IF THEY OBSERVE THAT THE REAGENT LID IS SHIFTING DOWNWARD WHILE IN THE OPEN POSITION.

Description of Event or Problem · 1

THE REAGENT LID OF A DIMENSION EXL 200 INSTRUMENT FELL ONTO AN OPERATOR, COMING INTO CONTACT WITH THEIR HEAD. THE OPERATOR SUSTAINED A LACERATION TO HER HEAD DURING THE INCIDENT. THE OPERATOR DID NOT REQUIRE MEDICAL TREATMENT BEYOND FIRST AID.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418270 DIMENSION EXL 200 CLINICAL CHEMISTRY ANALYZER JJE SIEMENS HEALTHCARE DIAGNOSTICS INC. DIMENSION EXL 200

Patients

Seq Age Sex Outcome Treatment
1