FDA Adverse Event Malfunction Summary report: N

DIMENSION VISTA 500

MDR report key: 3942479 · Received July 17, 2014

Report

Report Number
1226181-2014-00378
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 26, 2014
Report Date
June 27, 2014
Manufacturer
SIEMENS HEALTHCARE DIAGNOSTICS INC
Product Code
JJE
PMA / PMN Number
K051087
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
WI, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

THE CUSTOMER CONTACTED THE SIEMENS TECHNICAL SOLUTIONS CENTER (TSC). THE CUSTOMER STATED THAT THE SAMPLE WHICH RESULTED FALSELY HIGH WAS VISIBLY HEMOLYSED. OTHER SAMPLES OBTAINED FROM THE PATIENT RESULTED AS EXPECTED UPON REPEAT TESTING ON THE SAME INSTRUMENT. IT WAS DISCOVERED THAT THE CUSTOMER WAS CENTRIFUGING PATIENT SAMPLES OUTSIDE OF THE TUBE MANUFACTURER SPECIFICATIONS. THE CAUSE OF THE DISCORDANT, FALSELY ELEVATED TROPONIN RESULTS IS SAMPLE INTEGRITY. THE CAUSE OF THE SAMPLE BEING CENTRIFUGED OUTSIDE OF TUBE VENDOR SPECIFICATIONS IS FAILURE TO FOLLOW INSTRUCTIONS. THE INSTRUMENT IS PERFORMING ACCORDING TO SPECIFICATIONS. NO FURTHER EVALUATION OF THE DEVICE IS REQUIRED.

Description of Event or Problem · 1

DISCORDANT, FALSELY ELEVATED TROPONIN RESULTS WERE OBTAINED ON ONE PATIENT SAMPLE ON A DIMENSION VISTA 500 INSTRUMENT. THE INITIAL DISCORDANT, FALSELY ELEVATED TROPONIN RESULT WAS REPORTED TO THE PHYSICIAN(S), WHO QUESTIONED IT AS IT DID NOT MATCH THE RESULT FROM A PREVIOUS DRAW. A NEW SAMPLE OBTAINED FROM THE PATIENT WAS TESTED ON THE SAME INSTRUMENT, RESULTING LOWER. THE SAMPLE WHICH THE DISCORDANT RESULT WAS OBTAINED ON WAS THEN REPEATED ON THE SAME INSTRUMENT, RESULTING LOWER THAN THE FIRST RESULT PRODUCED BUT STILL FALSELY ELEVATED . THE CORRECTED RESULT FROM THE NEW SAMPLE WAS REPORTED TO THE PHYSICIAN(S). THERE ARE NO REPORTS OF PATIENT INTERVENTION OR ADVERSE HEALTH CONSEQUENCES DUE TO THE DISCORDANT, FALSELY ELEVATED TROPONIN RESULT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418566 DIMENSION VISTA 500 CLINICAL CHEMISTRY ANALYZER JJE SIEMENS HEALTHCARE DIAGNOSTICS INC DIMENSION VISTA 500

Patients

Seq Age Sex Outcome Treatment
1