PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2014-87578
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 17, 2014
- Report Date
- June 17, 2014
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WV
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.
THE INSULIN PUMP RECEIVED WITH UNRESPONSIVE ESC BUTTON DUE TO CORRODED KEYPAD TRACES. NO BUTTON ERROR ALARMS NOTED. NO TRACES OF MOISTURE WERE NOTED AT ELECTRONIC OR MOTOR ASSEMBLIES PER VISUAL INSPECTION. UNABLE TO PERFORM PRIME TEST OR DISPLACEMENT TEST DUE TO BUTTON ANOMALY. THE INSULIN PUMP RECEIVED WITH CRACKED CASE AT DISPLAY WINDOW CORNERS, CRACKED BATTERY TUBE THREADS, MINOR SCRATCHES ON DISPLAY WINDOW AND A LOOSE END CAP STICKER.
IT WAS REPORTED THAT THE INSULIN PUMP ALARMED BUTTON ERROR. THE CUSTOMER'S BLOOD GLUCOSE WAS 190 MG/DL. THE CUSTOMER STATED THAT THE DEVICE MIGHT HAVE GOTTEN EXPOSED TO SWEAT AT THE GYM. THE CUSTOMER ALSO STATED THAT THERE WAS A CRACK NEAR THE BATTERY CAP. THE DISCONTINUATION OF THE DEVICE WAS ADVISED. NOTHING FURTHER WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418988 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-723NAH |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 23 YR |