ULTRACISION HARMONIC ACE
Report
- Report Number
- 3005075853-2014-05062
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- July 10, 2014
- Report Date
- July 11, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- LFL
- PMA / PMN Number
- K042777
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). WHEN ADDITIONAL INFORMATION IS RECEIVED AND/OR THE DEVICE ANALYSIS HAS BEEN COMPLETED, A SUPPLEMENTAL MEDWATCH WILL BE SENT.
(B)(4). ADDITIONAL INFORMATION: DEVICE WAS REPROCESSED THE BLADE WAS NOTED TO HAVE DAMAGE IN THE FORM OF SCUFF/SCRATCHES AS IF CLEANED WITH A METAL PAD. IN ADDITION IT WAS NOTED THAT THE TISSUE PAD DETACHED AND A 4-40 STUD WAS BROKEN AND STUCK ON THE DEVICE. DUE TO THE BLADE DAMAGE ON THE INSTRUMENT WHICH IS EVIDENCE OF POSSIBLE REUSE/REPROCESSING, WE ARE UNABLE TO DETERMINE HOW THIS CONDITION IMPACTED THE PERFORMANCE OF THE DEVICE AND THEREFORE CANNOT CONCLUDE ROOT CAUSE. OUR MANUFACTURING, STERILIZATION, PACKAGING, AND SHIPMENT PROCESSES DO NOT INTRODUCE BLADE DAMAGE OF THIS TYPE TO THE DEVICE. THE DEVICE WAS DISASSEMBLED TO INSPECT INTERNAL COMPONENTS. CORROSION WAS FOUND AT THE HAND ACTIVATION DOMES. DUE TO THE MOISTURE DISCOVERED ON THE INSTRUMENT WHICH IS EVIDENCE OF POSSIBLE REUSE/REPROCESSING, WE ARE UNABLE TO DETERMINE HOW THIS CONDITION IMPACTED THE PERFORMANCE OF THE DEVICE AND THEREFORE CANNOT CONCLUDE ROOT CAUSE. OUR MANUFACTURING, STERILIZATION, PACKAGING, AND SHIPMENT PROCESSES DO NOT INTRODUCE CORROSION / MOISTURE TO THE DEVICE.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE THE TISSUE PAD FELL OFF. THE FALLEN PIECE WAS RETRIEVED BY FORCEPS. CHANGED TO ANOTHER ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT. ONE DEVICE WILL BE RETURNING.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418974 | ULTRACISION HARMONIC ACE | INSTRUMENT, ULTRASONIC SURGICAL | LFL | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | GENERATOR AND HANDPIECE |