FDA Adverse Event
Malfunction
Summary report: N
FLOGARD
MDR report key: 3942300
·
Received July 17, 2014
Report
- Report Number
- 1416980-2014-23078
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MX
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE BY A FIELD SERVICE TECHNICIAN. THE SAMPLE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED CONDITION OF A BATTERY LOW WAS VERIFIED IN THE ALARM LOG REVIEW AND DURING BATTERY TESTING. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO AN INOPERATIVE MAIN BATTERY. TO CORRECT THE ISSUE THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.
Description of Event or Problem · 1
IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED A BATTERY LOW ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.