FDA Adverse Event Malfunction Summary report: N

FLOGARD

MDR report key: 3942300 · Received July 17, 2014

Report

Report Number
1416980-2014-23078
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
BAXTER HEALTHCARE - SINGAPORE
Product Code
FRN
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MX
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE DEVICE WAS SERVICED AT THE CUSTOMER SITE BY A FIELD SERVICE TECHNICIAN. THE SAMPLE WAS VISUALLY INSPECTED AND FUNCTIONALLY TESTED. THE REPORTED CONDITION OF A BATTERY LOW WAS VERIFIED IN THE ALARM LOG REVIEW AND DURING BATTERY TESTING. THE CAUSE OF THE REPORTED CONDITION WAS DUE TO AN INOPERATIVE MAIN BATTERY. TO CORRECT THE ISSUE THE BATTERY WAS REPLACED. THE DEVICE WAS SERVICED AND RESTORED TO A GOOD WORKING CONDITION. IF ANY ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A FOLLOW UP MDR WILL BE SENT.

Description of Event or Problem · 1

IT WAS REPORTED THAT A FLOGARD INFUSION PUMP EXPERIENCED A BATTERY LOW ALARM. THERE WAS NO PATIENT INVOLVEMENT. ADDITIONAL INFORMATION WAS REQUESTED AND IS NOT AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419383 FLOGARD PUMP, INFUSION FRN BAXTER HEALTHCARE - SINGAPORE

Patients

Seq Age Sex Outcome Treatment
1