HOMECHOICE AUTOMATED PD SET WITH CASSETTE

MDR report key: 3942274 · Received July 17, 2014

Report

Report Number
1416980-2014-23067
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
FKX
PMA / PMN Number
K102936
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Additional Manufacturer Narrative · 1

(B)(4). AN ALARM INDICATIVE OF A POTENTIAL MALFUNCTION OF THE DISPOSABLE CASSETTE WAS IDENTIFIED. AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SYSTEM ERROR 2240 (AIR IN LINE/SET) ALARM OCCURRED ON A HOME CHOICE (HC) DEVICE DURING DRAIN ONE OF FOUR. THE HP DISCONNECTED, CYCLED THE POWER AND WENT THROUGH THE SYSTEM SETUP AGAIN AND THE HC BEGAN TO PRIME. THE TECHNICAL SERVICE REPRESENTATIVE (TSR) HAD THE HP PRESS STOP AND EXPLAINED ALARMS AND THAT THE HP WILL NEED TO START OVER WITH NEW SUPPLIES. THE TSR REVIEWED THE PROPER PROCEDURES PER USER MANUAL WITH THE HP. THERE WAS NO PATIENT INJURY OR ADVERSE EVENT ASSOCIATED WITH THIS REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419340 HOMECHOICE AUTOMATED PD SET WITH CASSETTE SYSTEM, PERITONEAL, AUTOMATIC DELIVERY FKX BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 HOMECHOICE