ARCHITECT C8000 SYSTEM
Report
- Report Number
- 1628664-2014-00141
- Event Type
- Injury
- Date Received
- July 17, 2014
- Date of Event
- June 25, 2014
- Report Date
- June 25, 2014
- Manufacturer
- ABBOTT MANUFACTURING INC
- Product Code
- JJE
- PMA / PMN Number
- K980367
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO
- Reporter Occupation
- PHYSICIAN
Narratives
ADDITIONAL INFORMATION WAS PROVIDED ON SEPTEMBER 11, 2014 REGARDING THE CURRENT STATE OF THE LABORATORY EMPLOYEE THAT SLIPPED AND FELL DUE TO WATER UNDERNEATH THE ARCHITECT C8000 ANALYER. THE EMPLOYEE WAS GIVEN 3 DAYS DISABILITY REST BY THE ACCOUNT'S OCCUPATIONAL RISK INSURANCE DEPARTMENT DUE TO HIS SPRAINED KNEE. THE EMPLOYEE RETURNED TO NORMAL WORK DUTIES AFTER THE 3 DAYS. THE INJURY HAS HEALED.
(B)(4). THE CUSTOMER WAS UNABLE TO IDENTIFY THE SOURCE OF THE WATER AND FIELD SERVICE WAS DISPATCHED. AN ABBOTT FIELD SERVICE ENGINEER (FSE) INSPECTED THE ANALYZER AND IDENTIFIED THAT THE TOP BELLOWS PUMP ELBOW FITTING (PART NUMBER 2-89191-02) FOR THE PROBE WASH PUMP WAS LOOSE AND LEAKING WATER. THE FSE TIGHTENED THE EXISTING ELBOW FITTING CONNECTION TO RESOLVE THE LEAKING. THE CAUSE OF THE LOOSE ELBOW FITTING CONNECTION COULD NOT BE DETERMINED. THE ELBOW FITTINGS ARE REMOVED WHEN THE CUSTOMER OR FSE PERFORMS MAINTENANCE OR TROUBLESHOOTING INVOLVING THE BELLOWS PUMP. NEITHER THE FSE NOR THE CUSTOMER HAD RECENTLY PERFORMED DOCUMENTED MAINTENANCE INVOLVING THE ELBOW FITTING CONNECTIONS. HOWEVER, A SERVICE TICKET OPENED TWO DAYS EARLIER INVOLVED VERIFICATION OF FLUIDIC COMPONENTS INCLUDING THE SYRINGES, CHECK VALVES, AND WASH CUPS AND VERIFICATION OF THE REAGENT AND SAMPLE PROBE CALIBRATIONS. NO ADVERSE TRENDS OR PRODUCT ISSUES WERE IDENTIFIED IN A 12 MONTH REVIEW OF BELLOWS PUMP ELBOW FITTING COMPLAINTS AND TRENDING DATA. THE CURRENT COMPLAINT IS THE SOLE TICKET DOCUMENTING A SLIP/FALL INJURY SUSTAINED DUE TO A LEAKING BELLOWS PUMP ELBOW FITTING OVER A THREE YEAR REVIEW PERIOD. THE ARCHITECT SYSTEM OPERATIONS MANUAL PROVIDES ADEQUATE LABELING WITH REGARD TO MAINTENANCE, COMPONENT REPLACEMENT, AND TROUBLESHOOTING THE REPORTED ISSUE. SECTION 9 STATES THE USER MAY NEED TO REPLACE CERTAIN SUPPLY AND PUMP COMPONENTS DUE TO NORMAL WEAR FROM DAILY OPERATIONS. INSTRUCTIONS INCLUDE REMOVING THE TOP OR SIDE ELBOW FITTING SECURING THE TUBING TO THE PUMP CONNECTION. WHEN REATTACHING THE ELBOW FITTING, THE USER IS INSTRUCTED TO FINGER-TIGHTEN THE FITTING AT THE PUMP CONNECTION. THERE IS INSUFFICIENT INFORMATION TO CONCLUDE A MALFUNCTION DID NOT OCCUR. THE CAUSE OF THE LOOSE ELBOW FITTING CONNECTION COULD NOT BE DETERMINED. THERE IS NO INDICATION OF A DEFICIENCY OF BELLOWS PUMP ELBOW FITTING, PART NUMBER 2-89191-02. THE EXISTING FITTING CONNECTION WAS TIGHTENED CORRECTLY AND REMAINS IN USE ON THE INSTRUMENT.
THE CUSTOMER STATED THAT SHE SAW WATER ON THE FLOOR UNDERNEATH THE LEFT SIDE OF THE ARCHITECT C8000 INSTRUMENT ON (B)(6) 2014. SHE WAS UNABLE TO IDENTIFY THE SOURCE OF THE WATER. SHE STATED THAT A LABORATORY EMPLOYEE FELL DUE TO THE WATER ON THE FLOOR. AN ABBOTT FIELD SERVICE ENGINEER INSPECTED THE C8000 ANALYZER AND REPAIRED A LEAK FROM A LOOSE PUMP FITTING. THE CUSTOMER WAS UNSURE HOW THE FITTING BECAME LOOSE. THE CUSTOMER STATED THAT THE EMPLOYEE WHO FELL WAS DIAGNOSED WITH A SPRAINED KNEE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419585 | ARCHITECT C8000 SYSTEM | AUTOMATED CHEMISTRY ANALYZER | JJE | ABBOTT MANUFACTURING INC |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Disability |