FDA Adverse Event Malfunction Summary report: N

ILAB ULTRASOUND IMAGING SYSTEM

MDR report key: 3942051 · Received July 17, 2014

Report

Report Number
2134265-2014-04290
Event Type
Malfunction
Date Received
July 17, 2014
Report Date
June 27, 2014
Manufacturer
BOSTON SCIENTIFIC - FREMONT (CE)
Product Code
OBJ
PMA / PMN Number
K051679
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IT
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Additional Manufacturer Narrative · 1

DEVICE EVALUATED BY MFR.: THE UNIT WAS RECEIVED IN GOOD CONDITION WITH NO VISIBLE DAMAGE OR DEFECTS OBSERVED. THE UNIT MEETS SPECIFICATIONS FOR MOTORDRIVE UNIT -5 QC FUNCTIONAL TEST. THE UNIT SUCCESSFULLY UNDERWENT A CONTINUOUS BURN-IN FOR OVERNIGHT. THE BATCH NUMBER IS UNKNOWN AND THE MANUFACTURING RECORDS FOR THE COMPLAINT DEVICE COULD NOT BE REVIEWED. THE ROOT CAUSE IS NOT CONFIRMED AS THERE WAS NO EVIDENCE OF THE ALLEGED ISSUE OR ANY ANOMALIES WHICH COULD HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. (B)(4).

Description of Event or Problem · 1

SAME CASE AS : 2134265-2014-04489 AND 2134265-2014-04488. IT WAS REPORTED THAT AUTOMATIC PULLBACK FAILURE OCCURRED. THE ILAB MOTOR DRIVE UNIT WAS USED IN CONJUNCTION WITH A BSC CORONARY CATHETER INTENDED TO VISUALIZE THE UNSPECIFIED LESION. DURING PROCEDURE, AUTOMATIC PULLBACK DID NOT WORK. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.

Description of Event or Problem · 1

SAME CASE AS : 2134265-2014-04489 AND 2134265-2014-04488. IT WAS REPORTED THAT AUTOMATIC PULLBACK FAILURE OCCURRED. THE ILAB MOTOR DRIVE UNIT WAS USED IN CONJUNCTION WITH A BSC CORONARY CATHETER INTENDED TO VISUALIZE THE UNSPECIFIED LESION. DURING PROCEDURE, AUTOMATIC PULLBACK DID NOT WORK. THE PROCEDURE WAS COMPLETED WITH THE SAME DEVICE. NO PATIENT COMPLICATIONS WERE REPORTED.