FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3942045 · Received March 5, 2014

Report

Report Number
1720753-2014-02017
Event Type
Malfunction
Date Received
March 5, 2014
Date of Event
February 10, 2014
Report Date
March 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

NO CONCLUSION CAN BE DRAWN AS CONCLUSIVE REPAIR INFO IS UNAVAILABLE AT THIS TIME AND NO ADD'L SERVICE INFO WAS PROVIDED.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE SYSTEM FAILED TO PERFORM FLUOROSCOPY. THERE IS NO REPORT OF INJURY OR DEATH ASSOCIATED WITH THE EVENT DESCRIBED IN THIS COMPLAINT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131354 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1