FDA Adverse Event Injury Summary report: N

ASPHERE M SPEC 12/14 36 -2

MDR report key: 3942040 · Received July 17, 2014

Report

Report Number
1818910-2014-23674
Event Type
Injury
Date Received
July 17, 2014
Date of Event
July 9, 2013
Report Date
March 25, 2015
Manufacturer
DEPUY ORTHOPAEDICS INC US
Product Code
JDI
PMA / PMN Number
K082585
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
WA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. REVIEW OF THE DEVICE HISTORY RECORDS AND/OR A LOT SPECIFIC COMPLAINT DATABASE SEARCH WAS NOT POSSIBLE AS THE PRODUCT AND LOT CODES REQUIRED WERE NOT PROVIDED. THE INVESTIGATION COULD NOT VERIFY OR IDENTIFY ANY PRODUCT CONTRIBUTION TO THE REPORTED EVENT WITH THE INFORMATION PROVIDED. BASED ON THE INABILITY TO IDENTIFY ROOT CAUSE, THE NEED FOR CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED AT THIS TIME. SHOULD THE PRODUCT AND/OR ADDITIONAL INFORMATION BE RECEIVED, THE INVESTIGATION WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Additional Manufacturer Narrative · 1

NO DEVICE ASSOCIATED WITH THIS REPORT WAS RECEIVED FOR EXAMINATION. A WORLDWIDE LOT SPECIFIC COMPLAINT DATABASE SEARCH, OR DEVICE HISTORY RECORD (DHR) REVIEW WAS NOT POSSIBLE BECAUSE THE REQUIRED LOT CODE(S) WAS NOT PROVIDED. BASED ON PREVIOUS INVESTIGATIONS THIS COMPLICATION OF JOINT REPLACEMENT IS UNLIKELY TO HAVE BEEN THE RESULT OF A DEVICE FAILING TO MEET REQUIRED SPECIFICATIONS. THE INFORMATION RECEIVED WILL BE RETAINED FOR POTENTIAL SERIES INVESTIGATIONS IF TRIGGERED BY TREND ANALYSIS, POST MARKET SURVEILLANCE, OR OTHER EVENTS WITHIN THE QUALITY SYSTEM. CORRECTIVE ACTION WAS NOT INDICATED. DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ADDITIONAL INFORMATION BE RECEIVED, THE INFORMATION WILL BE REVIEWED AND THE INVESTIGATION WILL BE RE-OPENED AS NECESSARY.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL MEDWATCH, A FOLLOW-UP MEDWATCH WILL BE FILED AS APPROPRIATE.

Description of Event or Problem · 1

PFS RECEIVED. AFTER REVIEW OF THE PFS FOR MDR REPORTABILITY, AN UNKNOWN STEM IS BEING ADDED FOR THE ALLEGED HIGH METAL IONS. NO LABS HAVE BEEN PROVIDED FOR THE ALLEGED HIGH METAL IONS.

Description of Event or Problem · 1

PATIENT WAS REVISED TO ADDRESS LINER MALPOSITIONING. UPDATE REC'D (B)(6) 2014 - LITIGATION PAPERS RECEIVED. IN ADDITION TO WHAT WAS PREVIOUSLY REPORTED, LITIGATION ALLEGES THE PATIENT SUFFERS FROM PAIN, DISCOMFORT, INFLAMMATION, AND LARGE AMOUNTS OF TOXIC COBALT CHROMIUM METAL ION PARTICLES TO BE RELEASED INTO THE BLOOD, TISSUE, AND BONE. DOI HAS BEEN PROVIDED. THE HEAD IS BEING ADDED TO ADDRESS METAL ON METAL ALLEGATIONS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418349 ASPHERE M SPEC 12/14 36 -2 ARTICULEZE HEAD (12/14 TAPER) : HIP METAL FEMORAL HEADS JDI DEPUY ORTHOPAEDICS INC US 3107614

Patients

Seq Age Sex Outcome Treatment
1 62 YR Required Intervention