ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2014-20483
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Report Date
- July 9, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- PATIENT
Narratives
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED, AND IT WAS EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2014 WITH THE FOLLOWING FINDINGS: THE DISPLAY SCREEN WAS FOUND TO BE DIM AND DISCOLORED. UNRELATED TO THE EVENT, THE BATTERY COMPARTMENT WAS OBSERVED TO BE CRACKED BELOW THE GRIP PAD.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING A DISPLAY (DIM/FADING/COLOR SPECTRUM) ISSUE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418684 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 29 YR |