FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 3942022 · Received March 5, 2014

Report

Report Number
1720753-2014-02020
Event Type
Malfunction
Date Received
March 5, 2014
Date of Event
February 12, 2014
Report Date
March 5, 2014
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ON SITE INVESTIGATION. THE BATTERY PACK WAS EVALUATED AND REPLACED. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM EXHIBITED 'PRE-CHARGE VOLTAGE ERRORS' UPON START UP. THIS ERROR IS LIKELY TO PREVENT THE SYSTEM FROM BOOTING TO A USABLE STATE. NO PT DEATH OR SERIOUS INJURY WAS REPORTED RELATED TO THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
131444 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1