NA
Report
- Report Number
- 2134070-2014-00114
- Event Type
- Malfunction
- Date Received
- June 3, 2014
- Date of Event
- May 13, 2014
- Report Date
- May 14, 2014
- Manufacturer
- STERILMED, INC.
- Product Code
- NMJ
- PMA / PMN Number
- K033579
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NY, US
- Reporter Occupation
- OTHER
Narratives
FINAL DEVICE INVESTIGATION FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION WITH NO USABLE CLIPS REMAINING. THE LAST CLIP, WHICH IS NOT DESIGNED TO BE USED, WAS ADVANCED PARTIALLY INTO THE JAW. THE JAW APPEARED TO BE IN ALIGNMENT AND THE CLIP RETAINER, PUSH FORK AND JAW CLEARANCE APPEARED TO BE ACCEPTABLE. UPON EVALUATION, THE LOCKING MECHANISM WAS FOUND TO BE ENGAGED AS INTENDED. NO LOT NUMBER WAS PROVIDED, SO THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. AS THE RETURNED HAD NO USABLE CLIPS, THE COMPLAINT COULD NOT BE CONFIRMED.
IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY THE CLIPS FROM THREE DEVICES DID NOT FORM PROPERLY AND WASHED OFF THE ARTERY. THEY DID NOT COMPLETELY CLOSE AND LEFT A LOOP AT THE BEND AREA. CLIPS HAD BEEN FIRED SUCCESSFULLY FROM ONE OR MORE OF THE DEVICES. THERE WAS NO TORQUING OR TWISTING OF THE DEVICES AT THE TIME OF FIRING AND THERE WAS NO UNEXPECTED RESISTANCE FELT WHILE FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THIS REPORT IS FOR THE THIRD DEVICE. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO ADDITIONAL INFORMATION WAS RECEIVED. A SUPPLEMENTAL REPORT WILL BE SENT IF ADDITIONAL INFORMATION BECOMES AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 323389 | NA | NMJ | STERILMED, INC. | ETHER320 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |