FDA Adverse Event Malfunction Summary report: N

NA

MDR report key: 3941870 · Received June 3, 2014

Report

Report Number
2134070-2014-00114
Event Type
Malfunction
Date Received
June 3, 2014
Date of Event
May 13, 2014
Report Date
May 14, 2014
Manufacturer
STERILMED, INC.
Product Code
NMJ
PMA / PMN Number
K033579
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FINAL DEVICE INVESTIGATION FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION WITH NO USABLE CLIPS REMAINING. THE LAST CLIP, WHICH IS NOT DESIGNED TO BE USED, WAS ADVANCED PARTIALLY INTO THE JAW. THE JAW APPEARED TO BE IN ALIGNMENT AND THE CLIP RETAINER, PUSH FORK AND JAW CLEARANCE APPEARED TO BE ACCEPTABLE. UPON EVALUATION, THE LOCKING MECHANISM WAS FOUND TO BE ENGAGED AS INTENDED. NO LOT NUMBER WAS PROVIDED, SO THE DEVICE HISTORY RECORD COULD NOT BE REVIEWED. AS THE RETURNED HAD NO USABLE CLIPS, THE COMPLAINT COULD NOT BE CONFIRMED.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY THE CLIPS FROM THREE DEVICES DID NOT FORM PROPERLY AND WASHED OFF THE ARTERY. THEY DID NOT COMPLETELY CLOSE AND LEFT A LOOP AT THE BEND AREA. CLIPS HAD BEEN FIRED SUCCESSFULLY FROM ONE OR MORE OF THE DEVICES. THERE WAS NO TORQUING OR TWISTING OF THE DEVICES AT THE TIME OF FIRING AND THERE WAS NO UNEXPECTED RESISTANCE FELT WHILE FIRING. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THIS REPORT IS FOR THE THIRD DEVICE. ADDITIONAL INFORMATION WAS REQUESTED, BUT NO ADDITIONAL INFORMATION WAS RECEIVED. A SUPPLEMENTAL REPORT WILL BE SENT IF ADDITIONAL INFORMATION BECOMES AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
323389 NA NMJ STERILMED, INC. ETHER320

Patients

Seq Age Sex Outcome Treatment
1