MARYLAND BIPOLAR FORCEPS INSTRUMENT
Report
- Report Number
- 2955842-2014-04345
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 18, 2014
- Report Date
- June 23, 2014
- Manufacturer
- INTUITIVE SURGICAL,INC.
- Product Code
- NAY
- PMA / PMN Number
- K050369
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PHYSICIAN
Narratives
THE INSTRUMENT WAS RETURNED AND EVALUATED. PER THE CUSTOMER REPORTED COMPLAINT, FAILURE ANALYSIS INVESTIGATION FOUND PITCH UP CABLE FRAYED AT THE DISTAL CLEVIS HUB. THE FRAYED STRANDS STUCK OUT AT THE INSTRUMENT'S WRIST. OTHER CABLES AT THE WRIST WERE NOT DAMAGED. NO OTHER DAMAGE WAS FOUND. THE CUSTOMER REPORTED COMPLAINT DOES NOT ITSELF CONSTITUTE A REPORTABLE EVENT; HOWEVER, THE FRAYED CABLE COULD LIKELY CAUSE OR CONTRIBUTE TO AN ADVERSE EVENT, IF THE MALFUNCTION WERE TO RECUR.
IT WAS REPORTED THAT DURING A DA VINCI PROSTATECTOMY PROCEDURE, IT WAS CONFIRMED THAT THE WIRE WAS BROKEN ON A MARYLAND BIPOLAR FORCEPS INSTRUMENT. USING BACKUP INSTRUMENT, THE PLANNED PROCEDURE WAS COMPLETED WITHOUT PROBLEMS. THERE WAS NO ALLEGATION OF HARM OR INJURY TO A PATIENT. NOTHING REPORTEDLY FELL INTO THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418458 | MARYLAND BIPOLAR FORCEPS INSTRUMENT | ENDOSCOPIC ELECTROSURGICAL INSTRUMENT | NAY | INTUITIVE SURGICAL,INC. | 420172-07 | M10131017 664 |