FDA Adverse Event Malfunction Summary report: N

RESERVOIR 1.8ML

MDR report key: 3941764 · Received July 17, 2014

Report

Report Number
3004209178-2014-87493
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 17, 2014
Report Date
June 17, 2014
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
FRN
PMA / PMN Number
K001828
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Description of Event or Problem · 1

IT WAS REPORTED BY THE CUSTOMER THAT THEIR INSULIN PUMP WAS ALARMING NO DELIVERY. CUSTOMER STATES THAT THEY RECEIVED THE ALARM WHILE PRIMING AND WAS ABLE TO TROUBLESHOOT THE DEVICE THEMSELVES. CUSTOMER CLEARED ALARM BY CHANGING RESERVOIR AND INFUSION SET. CUSTOMER'S BLOOD GLUCOSE LEVEL WAS 432 MG/DL AT TIME OF REPORTING. NOTHING FURTHER REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419724 RESERVOIR 1.8ML RESERVOIR FRN MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-326A

Patients

Seq Age Sex Outcome Treatment
1 7 YR