FDA Adverse Event Malfunction Summary report: N

OPTICROSS?

MDR report key: 3941750 · Received July 17, 2014

Report

Report Number
2134265-2014-04343
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
May 26, 2014
Report Date
June 30, 2014
Manufacturer
BOSTON SCIENTIFIC - FREMONT (SUD)
Product Code
OBJ
PMA / PMN Number
K123621
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AGE AT TIME OF EVENT: UNDER 18 YEARS. DEVICE EVALUATED BY MANUFACTURER: THE COMPLAINT DEVICE WAS RECEIVED FOR EVALUATION. EVALUATION OF THE RETURNED DEVICE REVEALED AN OPEN HOLE AT THE SHEATH LAP JOINT SECTION OF THE DEVICE. FLUID WAS LEAKING FROM THE OPEN HOLE AT THE SHEATH LAP JOINT ASSEMBLY WHEN THE CATHETER WAS FLUSHED. THE TELESCOPE ASSEMBLY WAS NOT ABLE TO PROPERLY PULL BACK, ADVANCE, OR RETRACT. SINCE THE TELESCOPE CANNOT ADVANCE THE TRANSDUCER DISTAL HOUSING (TDH) TO THE MOST DISTAL POSITION, THE DISTANCE FROM THE DISTAL END OF THE TRANSDUCER HOUSING TO THE TIP OF THE CATHETER WAS NOT MEASURED. DURING IMAGE CHARACTERIZATION TESTING IN THE ROLLER COASTER MODEL, A GOOD SQUARE IMAGE APPEARED IN THE SYSTEM. NO IMAGING CORE WINDUP WAS FOUND WITHIN THE TELESCOPE SECTION OF THE DEVICE. NO OTHER ISSUES OR DEFECTS WERE OBSERVED DURING PRODUCT ANALYSIS OF THE RETURNED DEVICE. THE MANUFACTURING BATCH RECORD REVIEW CONFIRMED THAT THE DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. THE ROOT CAUSE IS NOT CONFIRMED AS THERE WAS NO EVIDENCE OF THE ALLEGED ISSUE OR ANY ANOMALIES WHICH COULD HAVE CONTRIBUTED TO THE REPORTED DIFFICULTY. (B)(4).

Description of Event or Problem · 1

REPORTABLE BASED ON DEVICE ANALYSIS COMPLETED ON (B)(4) 2014. IT WAS REPORTED THAT LOSS OF IMAGE OCCURRED. THE TARGET LESION WAS LOCATED IN THE SEVERELY CALCIFIED PROXIMAL RIGHT CORONARY ARTERY. DURING PERCUTANEOUS CORONARY INTERVENTION (PCI), AN OPTICROSS IMAGING CATHETER WAS USED TO VISUALIZE THE VESSEL HOWEVER, IMAGE DISAPPEARED DURING SECOND PULLBACK. THE PROCEDURE WAS COMPLETED WITH ANOTHER WITH SAME DEVICE. NO PATIENT COMPLICATIONS REPORTED AND THE PATIENT'S CONDITION IS GOOD. HOWEVER, DEVICE ANALYSIS REVEALED AN OPEN HOLE AT THE SHEATH LAP JOINT SECTION OF THE DEVICE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
418968 OPTICROSS? CATHETER, ULTRASOUND, INTRAVASCULAR OBJ BOSTON SCIENTIFIC - FREMONT (SUD) H749518080 16770581

Patients

Seq Age Sex Outcome Treatment
1