HOMECHOICE AUTOMATED PD SET WITH CASSETTE
Report
- Report Number
- 1416980-2014-23014
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 22, 2014
- Report Date
- June 22, 2014
- Manufacturer
- BAXTER HEALTHCARE - MOUNTAIN HOME
- Product Code
- FKX
- PMA / PMN Number
- K102936
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). AS THE CASSETTE WAS NOT RETURNED AND THE LOT NUMBER WAS UNKNOWN, A DEVICE ANALYSIS COULD NOT BE COMPLETED. HOWEVER, A CARE GIVER (CG) REPORTED THAT THE HOME PATIENT (HP) CONNECTED TO THE PATIENT LINE BEFORE THE LINE WAS PROPERLY PRIMED. THE HOMECHOICE AND HOMECHOICE PRO APD SYSTEMS PATIENT AT-HOME GUIDE, WHICH IS SHIPPED WITH EVERY HOMECHOICE DEVICE PROVIDES STEP-BY-STEP INSTRUCTIONS FOR PROPERLY PRIMING THE DISPOSABLE SET AND SAYS TO VERIFY THAT THE PATIENT LINE IS PROPERLY PRIMED. IT WARNS THE USER NOT TO CONNECT TO THE PATIENT LINE UNLESS THE FLUID LEVEL IS AT OR NEAR THE CONNECTOR AT THE END OF THE DISPOSABLE SET PATIENT LINE. A FORMAL REVIEW OF THE LABEL FOR THE PRODUCT FAMILY WILL BE CONDUCTED. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED.
WHILE TROUBLESHOOTING FOR A CHECK PATIENT LINE ALARM ON A HOMECHOICE (HC) DEVICE, A TECHNICAL SERVICE REPRESENTATIVE (TSR) WAS INFORMED THAT THERE WERE GAPS OF AIR IN THE PATIENT LINE. THE AIR WAS OBSERVED DURING THE INITIAL DRAIN PHASE OF AUTOMATED PERITONEAL DIALYSIS (APD) THERAPY. THE HOME PATIENT (HP) WAS CONNECTED WHEN THE AIR WAS NOTED. TROUBLESHOOTING FOR THE OBSERVED AIR REVEALED THAT THE HP FAILED TO ALLOW THE PATIENT LINE TO PRIME PROPERLY PRIOR TO CONNECTING TO THE DEVICE. THE TSR INSTRUCTED A CARE GIVER (CG) TO RESTART THERAPY USING NEW SUPPLIES. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION INDICATED AT THE TIME OF THE REPORT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419862 | HOMECHOICE AUTOMATED PD SET WITH CASSETTE | SYSTEM, PERITONEAL, AUTOMATIC DELIVERY | FKX | BAXTER HEALTHCARE - MOUNTAIN HOME |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 3 YR | HOMECHOICE |