FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3941547 · Received April 17, 2014

Report

Report Number
3008642652-2014-01043
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 20, 2014
Report Date
April 2, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
BIOMEDICAL ENGINEER

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. AS RECEIVED, THE TRUNK CABLE CONNECTOR WOULD NOT SECURELY CONNECT WITH THE MONITOR. THE CAUSE FOR THE INABILITY TO PROPERLY CONNECT WITH THE MONITOR IS A DEFECTIVE TRUNK CABLE CONNECTOR. THE ROOT CAUSE HAS NOT BEEN POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE CONNECTOR. THE LAST PATIENT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Description of Event or Problem · 1

A REVIEW OF SERVICE DATA HAS DETECTED A REPORTABLE PROBLEM. DURING SERVICING, ELECTRODE BELT SN (B)(4) WOULD NOT SECURELY LATCH TO A MONITOR. THE LAST PATIENT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236094 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 NA