ROTATING MULTIPLE CLIP APPLIER
Report
- Report Number
- 3005075853-2014-05040
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- July 2, 2014
- Report Date
- July 4, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). ADDITIONAL INFORMATION: THE ANALYSIS RESULTS OF THE ER420 DEVICE FOUND THAT IT WAS RECEIVED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT THE DEVICE WAS FUNCTIONALLY EVALUATED. UPON FIRING OF THE DEVICE, THE REMAINING CLIPS WERE EJECTED AND THEN, IT LOCKED OUT AS INTENDED. IT COULD NOT BE DETERMINED WHAT MAY HAVE CAUSE THE FOUND CONDITION. THE BATCH RECORD WAS REVIEWED AND NO ANOMALIES WERE NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THERE WERE UNFORMED CLIPS DEPLOYED AFTER FIRING. CHANGED TO ANOTHER ONE BUT STILL HAD THE SAME ISSUE. CHANGED THE THIRD ONE TO COMPLETE THE PROCEDURE. NO ADVERSE EVENT ON THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418072 | ROTATING MULTIPLE CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | NA | K4CX6X |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |