ANIMAS VIBE
Report
- Report Number
- 2531779-2014-20416
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Report Date
- June 26, 2014
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- PATIENT
Narratives
DEVICE EVALUATION: THE DEVICE HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON 08/15/2014 WITH THE FOLLOWING FINDINGS: THE COMPLAINT OF A DIM, FADED DISPLAY COULD NOT BE INVESTIGATED BECAUSE THE PUMP WOULD NOT POWER ON. THERE WAS EVIDENCE OF MOISTURE FOUND INSIDE THE PUMP, AND CORROSION VISIBLE IN THE BATTERY COMPARTMENT. LEAK TESTING REVEALED A BATTERY COMPARTMENT LEAK. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
THE PUMP HAS NOT BEEN RETURNED TO ANIMAS FOR EVALUATION. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.
ON (B)(6) 2014, THE REPORTER CONTACTED ANIMAS, ALLEGING THAT THE DISPLAY WAS DIM/FADED/DISCOLORED. REPORTEDLY, ADJUSTING THE CONTRAST DID NOT RESOLVE THE ISSUE BUT THE SCREEN WAS STILL LEGIBLE. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418082 | ANIMAS VIBE | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR |