FDA Adverse Event Malfunction Summary report: N

ESPRIT VENTILATOR

MDR report key: 3941389 · Received July 17, 2014

Report

Report Number
2031642-2014-00693
Event Type
Malfunction
Date Received
July 17, 2014
Report Date
June 24, 2014
Manufacturer
RESPIRONICS CALIFORNIA, INC.
Product Code
CBK
PMA / PMN Number
K981072
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
ND, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

THE CUSTOMER REPORTED THE VENTILATOR POWER SUPPLY FAILED. THE CUSTOMER REPORTED THE DEVICE WAS NOT IN USE ON A PATIENT THEREFORE THERE WAS NO PATIENT INVOLVEMENT OR HARM. THE MANUFACTURERS FIELD SERVICE ENGINEER (FSE) CONFIRMED THE REPORTED POWER SUPPLY FAILURE. DURING EVALUATION, THE FSE OBSERVED THE MAINS POWER LED INDICATOR WAS FLASHING AND AN ODOR OF OVERHEATING FROM THE POWER SUPPLY AREA. THE FSE REPLACED THE POWER SUPPLY TO ADDRESS THE REPORTED PROBLEM. APPLICABLE FINAL TESTING WAS PERFORMED PER OPERATING SPECIFICATIONS.

Description of Event or Problem · 1

FACTORY ANALYSIS OF THE RETURNED POWER SUPPLY REVEALED A BRIDGE RECTIFIER SHORT THAT RESULTED IN THE REPORTED POWER ISSUE. EVALUATION OF THE COMPONENT NOTED HEAT RELATED DAMAGE. THE FAILURE AS NOTED MAY RESULT IN A LOSS OF AC POWER DURING NORMAL VENTILATION OPERATION. IF A FAILURE OF THE POWER SUPPLY OCCURS DURING USE AND THE VENTILATOR SHUTS DOWN DUE TO A LOSS OF AC POWER, AN AUDIBLE POWER FAIL ALARM WILL ACTIVATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
417940 ESPRIT VENTILATOR VENTILATOR, CONTINUOUS CBK RESPIRONICS CALIFORNIA, INC. V1000 NA