PROGRAMMING SOFTWARE
Report
- Report Number
- 1644487-2014-01812
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- CYBERONICS, INC.
- Product Code
- LYJ
- PMA / PMN Number
- P970003
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- PHYSICIAN
Narratives
ANALYSIS OF THE RETURNED HANDHELD DEVICE WAS COMPLETED. NO ANOMALIES ASSOCIATED WITH THE HANDHELD PERFORMANCE WERE NOTED DURING TESTING USING THE AC ADAPTER OR THE MAIN BATTERY WITH A FULL CHARGE. THE HANDHELD PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS. NO ANOMALIES ASSOCIATED WITH FLASHCARD SOFTWARE OR DATABASES WERE IDENTIFIED DURING THE FLASHCARD ANALYSIS. THE FLASHCARD AND SOFTWARE PERFORMED ACCORDING TO FUNCTIONAL SPECIFICATIONS.
IT WAS REPORTED THAT THE PHYSICIAN KEPT THE CHARGER AND THAT THE CHARGER WORKS WITH WITH OTHER HANDHELDS.
IT WAS REPORTED THAT THE PHYSICIAN'S HANDHELD WILL NOT HOLD A CHARGE. THE PHYSICIAN WAS PROVIDED A NEW PROGRAMMING COMPUTER AND THE HANDHELD WAS RETURNED FOR ANALYSIS. ANALYSIS IS UNDERWAY, BUT HAS NOT BEEN COMPLETED TO DATE. ATTEMPTS TO OBTAIN ADDITIONAL RELEVANT INFORMATION HAVE BEEN UNSUCCESSFUL TO DATE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418053 | PROGRAMMING SOFTWARE | PROGRAMMING COMPUTER | LYJ | CYBERONICS, INC. | MODEL 250 | 1075663 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |