Manufacturer
ZOLL LIFECOR CORP
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVAL SUMMARY: DEVICE EVAL OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (CHECK BELT ALARMS) HAS BEEN CONFIRMED. UPON INVESTIGATION THE CABLE CONNECTING THE DISTRIBUTION NODE TO THE REAR THERAPY ELECTRODES WAS PULLED FROM THE STRAIN RELIEF, DAMAGING WIRES. THE ROOT CAUSE OF THE STRAINED CABLE COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE STRAINED CABLE. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HE WAS RECEIVING FREQUENT CHECK BELT ALARMS. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.