FDA Adverse Event Malfunction Summary report: N

MAESTRO MEDIUM FIXED DURAGUARD

MDR report key: 3941193 · Received July 17, 2014

Report

Report Number
0001811755-2014-02559
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 23, 2014
Report Date
June 23, 2014
Manufacturer
STRYKER INSTRUMENTS-KALAMAZOO
Product Code
HBB
PMA / PMN Number
K041754
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE REPORTED EVENT OF A BENT TIP ON THE DURAGUARD WAS CONFIRMED BY A STRYKER SERVICE TECHNICIAN THROUGH VISUAL INSPECTION OF A BENT FOOT ON THE DURAGUARD. BASED ON THE RISK DOCUMENTATION A BENT DURAGUARD CAN OCCUR WHEN EXCESSIVE SIDE LOAD IS APPLIED TO THE FEATURE. DEVICE IS SCRAPPED BY STRYKER.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE TIP OF THE DURAGUARD WAS BENT ALLOWING FOR A CONDITION WHERE THE BUR/CUTTING ACCESSORY HAS THE POTENTIAL TO DAMAGE THE BLOOD VESSEL OR MEMBRANE. IT WAS FOUND DURING SET UP AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419745 MAESTRO MEDIUM FIXED DURAGUARD MOTOR, DRILL, PNEUMATIC HBB STRYKER INSTRUMENTS-KALAMAZOO 13247

Patients

Seq Age Sex Outcome Treatment
1