FDA Adverse Event
Malfunction
Summary report: N
MAESTRO MEDIUM FIXED DURAGUARD
MDR report key: 3941193
·
Received July 17, 2014
Report
- Report Number
- 0001811755-2014-02559
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HBB
- PMA / PMN Number
- K041754
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED EVENT OF A BENT TIP ON THE DURAGUARD WAS CONFIRMED BY A STRYKER SERVICE TECHNICIAN THROUGH VISUAL INSPECTION OF A BENT FOOT ON THE DURAGUARD. BASED ON THE RISK DOCUMENTATION A BENT DURAGUARD CAN OCCUR WHEN EXCESSIVE SIDE LOAD IS APPLIED TO THE FEATURE. DEVICE IS SCRAPPED BY STRYKER.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE TIP OF THE DURAGUARD WAS BENT ALLOWING FOR A CONDITION WHERE THE BUR/CUTTING ACCESSORY HAS THE POTENTIAL TO DAMAGE THE BLOOD VESSEL OR MEMBRANE. IT WAS FOUND DURING SET UP AT THE USER FACILITY. THE PROCEDURE WAS COMPLETED SUCCESSFULLY WITH NO DELAY. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419745 | MAESTRO MEDIUM FIXED DURAGUARD | MOTOR, DRILL, PNEUMATIC | HBB | STRYKER INSTRUMENTS-KALAMAZOO | 13247 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |