SYSTEM 6 SAGITTAL SAW
MDR report key: 3941191
·
Received July 17, 2014
Report
- Report Number
- 0001811755-2014-02558
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- June 23, 2014
- Report Date
- June 23, 2014
- Manufacturer
- STRYKER INSTRUMENTS-KALAMAZOO
- Product Code
- HWE
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MI, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE REPORTED DISASSEMBLY WAS CONFIRMED BY A MANUFACTURER REPAIR TECHNICIAN THROUGH VISUAL INSPECTION. UPON DISASSEMBLY, IT WAS FOUND THAT THE TRIGGER HOUSING WAS CRACKED AT THE SCREW BOSS, WHICH CAN LEAD TO THE REPORTED EVENT AND CAN BE CAUSED BY A MATERIAL FATIGUE ISSUE OR IMPACT.
Description of Event or Problem · 1
IT WAS REPORTED THAT DURING INCOMING INSPECTION AT THE MANUFACTURER FACILITY THE STRAIGHT PIN OF THE DEVICE DISASSEMBLED. NO MEDICAL INTERVENTION AND NO ADVERSE CONSEQUENCES WERE REPORTED WITH THIS EVENT. AS THIS EVENT OCCURRED DURING TESTING AT THE MANUFACTURER FACILITY, THERE WAS NO PATIENT INVOLVEMENT AND NO DELAY TO A SURGICAL PROCEDURE.