FDA Adverse Event Malfunction Summary report: N

PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP

MDR report key: 3941142 · Received July 17, 2014

Report

Report Number
2032227-2014-04308
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
June 11, 2014
Report Date
June 16, 2014
Manufacturer
MEDTRONIC MINIMED
Product Code
OYC
PMA / PMN Number
P980022
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN
Reporter Occupation
PATIENT

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT. THE DEVICE HAS BEEN RETURNED, BUT NOT YET EVALUATED. FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED. NO CONCLUSION CAN BE DRAWN AT THIS TIME.

Additional Manufacturer Narrative · 1

INSULIN PUMP RECEIVED WITH BLANK DISPLAY DUE TO MOISTURE DAMAGE ELECTRONIC ASSEMBLY AND MOTOR. INSULIN PUMP UNABLE TO VERIFY BUTTON ERROR ALARM, NO BUTTON RESPONSE ANOMALY, REWIND ANOMALY DUE TO BLANK DISPLAY. INSULIN PUMP HAD MINOR SCRATCHES ON DISPLAY WINDOW AND CRACKED RESERVOIR TUBE LIP.

Description of Event or Problem · 1

CUSTOMER REPORTED THAT HER INSULIN PUMP WAS MALFUNCTIONING, IT WAS ALARMING BUTTON ERROR, UNABLE TO REWIND AND THE KEYPAD WAS UNRESPONSIVE. CUSTOMER STATED THAT LAST WEEK SHE GOT IN THE POOL WITH THE INSULIN PUMP. NOTHING FURTHER WAS REPORTED.