FDA Adverse Event Malfunction Summary report: N

FOLFUSOR

MDR report key: 3941129 · Received July 17, 2014

Report

Report Number
1416980-2014-22974
Event Type
Malfunction
Date Received
July 17, 2014
Report Date
June 23, 2014
Manufacturer
BAXTER HEALTHCARE - IRVINE
Product Code
MEB
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BE
Reporter Occupation
PHARMACIST

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE ACTUAL DEVICE WAS EVALUATED. VISUAL INSPECTION OF THE DEVICE DID NOT REVEAL ANY ABNORMALITIES THAT COULD HAVE CONTRIBUTED TO THE REPORTED CONDITION. A FUNCTIONAL FLOW RATE TESTING WAS PERFORMED AND THE DEVICE WAS FOUND WITHIN SPECIFICATION. THE CAUSE OF THE CONDITION COULD NOT BE DETERMINED. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.

Description of Event or Problem · 1

IT WAS REPORTED THAT A SMALL VOLUME FOLFUSOR UNDERINFUSED. THE REPORTER STATED THAT THE DEVICE WAS NOT COMPLETELY EMPTY WHEN THE PATIENT RETURNED TO THE HOSPITAL AFTER THE EXPECTED INFUSION TIME. THERE WAS NO PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419398 FOLFUSOR PUMP, INFUSION, ELASTOMERIC MEB BAXTER HEALTHCARE - IRVINE 13G057

Patients

Seq Age Sex Outcome Treatment
1