FDA Adverse Event Malfunction Summary report: N

ILS 29MM CURVED

MDR report key: 3941128 · Received July 17, 2014

Report

Report Number
3005075853-2014-05030
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
March 3, 2014
Report Date
May 28, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K983536
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
BR
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INCOMPLETE FIRING. THE ANALYSIS RESULTS FOUND THAT THE CDH29A DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (THE FIRING HANDLE REACHES ITS STOPPING POINT, AND THE FIRING TRIGGER IS PARALLEL TO THE INSTRUMENT HANDLE) STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. PLEASE REFERENCE THE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE PRODUCT MALFUNCTIONED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
420002 ILS 29MM CURVED STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA K4DC4N

Patients

Seq Age Sex Outcome Treatment
1