ILS 29MM CURVED
Report
- Report Number
- 3005075853-2014-05030
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- March 3, 2014
- Report Date
- May 28, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K983536
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- BR
- Reporter Occupation
- OTHER
Narratives
(B)(4). INCOMPLETE FIRING. THE ANALYSIS RESULTS FOUND THAT THE CDH29A DEVICE ARRIVED IN GOOD VISUAL CONDITION. THE BREAKAWAY WASHER UNCUT AND INDENTED, INDICATING THAT THE DEVICE HAD NOT BEEN FIRED THROUGH A FULL FIRING STROKE OR POSSIBLY THAT THE ORANGE INDICATOR WAS NOT FULLY INTO THE SAFE GREEN FIRING RANGE. IT SHOULD BE NOTED THAT BEFORE FIRING THE DEVICE THE ORANGE INDICATOR SHOULD BE FULLY WITHIN THE GREEN RANGE OF THE GAP SETTING SCALE. IN ADDITION IT SHOULD BE NOTED THAT IF THE FIRING SEQUENCE IS NOT COMPLETE (THE FIRING HANDLE REACHES ITS STOPPING POINT, AND THE FIRING TRIGGER IS PARALLEL TO THE INSTRUMENT HANDLE) STAPLES COULD BE PARTIALLY DEPLOYED WITHOUT CUTTING THE WASHER. PLEASE REFERENCE THE INSTRUCTIONS FOR USE FOR ADDITIONAL INFORMATION. THE DEVICE WAS RELOADED WITH STAPLES AND TESTED FOR FUNCTIONALITY WITH A TEST WASHER; THE DEVICE FORMED ALL THE STAPLES, AS WELL AS COMPLETELY CUT THE TEST MEDIA AND THE BREAKAWAY WASHER WITHOUT INCIDENT. THE STAPLE LINE WAS COMPLETE AND THE STAPLES WERE NOTED TO HAVE THE PROPER B-FORMED SHAPE. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE PRODUCT MALFUNCTIONED. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE. THERE WERE NO ADVERSE CONSEQUENCES FOR THE PATIENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 420002 | ILS 29MM CURVED | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | K4DC4N |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |