CARE ASSIST BED
Report
- Report Number
- 9615739-2014-00170
- Event Type
- Malfunction
- Date Received
- April 23, 2014
- Date of Event
- March 26, 2014
- Report Date
- March 26, 2014
- Manufacturer
- HILL-ROM PLUVIGNER
- Product Code
- FNL
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICIAN FOUND ALL 4 CASTER NOT HOLDING MOST LIKELY DUE TO NORMAL WEAR AND TEAR. PER THE HILL-ROM SERVICE MANUAL THE BED SHOULD BE SUBJECT TO AN EFFECTIVE MAINTENANCE PROGRAM. AN ANNUAL SERVICE OF THE BED IS ADVISED IN ORDER TO MAINTAIN ITS CHARACTERISTICS AND PERFORMANCE. BRAKE CASTERS SHOULD BE CHECKED FOR CUTS, WEAR AND QUALITY OF TREAD, ETC. AND REPLACED WHEN NECESSARY. CHECK THE BRAKES TO SEE WHETHER THE BED MOVES WHEN THE BRAKE PEDALS ARE PRESSED AND REPAIR AS NECESSARY. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THIS BED IN 2012 AND 2014. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE OF THIS BED. THE TECHNICIAN REPLACED ALL 4 CASTERS, CASTER SUPPORTS, AND BOTH MAIN BEARINGS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES ARE NOT HOLDING. THE BED WAS LOCATED AT THE ACCOUNT IN MAINTENANCE. THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS (B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 245170 | CARE ASSIST BED | A/C POWERED ADJUSTABLE HOSPITAL BED | FNL | HILL-ROM PLUVIGNER | 1170 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |