75MM SELECTABLE NEW TLC
Report
- Report Number
- 3005075853-2014-05029
- Event Type
- Malfunction
- Date Received
- July 17, 2014
- Date of Event
- May 20, 2014
- Report Date
- May 27, 2014
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- GDW
- PMA / PMN Number
- K092577
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). FIRING MECHANISM. THE ANALYSIS RESULTS FOUND THAT THE NTLC75 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING KNOB WAS DETACHED AND THE SLIP BLOCK ASSEMBLY WAS NOTED TO BE DAMAGED. THE DEVICE WAS RECEIVED WITH ONE RELOAD PRESENT. THE RELOAD WAS RETURNED WITH THE PROXIMAL 57 DRIVERS UP WITHOUT STAPLES, AND THE REMAINING DRIVERS DOWN WITH STAPLES PRESENT; THE SWING TAB IN THE LOCKED POSITION, WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. THE DAMAGE TO THE FIRING KNOB AND SLIP BLOCK ASSEMBLY IS CONSISTENT WITH HIGH (OUTSIDE INDICATED USE) STAPLE FORMING FORCES; HOWEVER THERE IS INSUFFICIENT EVIDENCE TO DETERMINE THE CAUSE OF THE HIGHER LOADS. IT SHOULD BE NOTED THAT THE CARTRIDGE RELOAD IS DESIGNED TO LOCKOUT, AS A SAFETY FEATURE, IF ANY STAPLES HAVE BEEN FIRED FROM THE CARTRIDGE RELOAD. IF ENOUGH FORCE IS APPLIED THE DEVICE COULD BE DAMAGED. IN ADDITION FAILURE TO COMPLETE THE STROKE MAY RESULT IN INCOMPLETE STAPLE LINE. FOR ADDITIONAL INFORMATION PLEASE REFER TO THE INSTRUCTIONS FOR USE. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TEST COULD BE PERFORMED WITH IT. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.
IT WAS REPORTED THAT DURING AN OPEN COLON PROCEDURE, ACCORDING TO THE SURGEON ON THE FOURTH FIRING THE STAPLER JAMMED. THE BLADE WOULD NOT MOVE FORWARD AND THEN THE HANDLE BROKE OFF. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 419000 | 75MM SELECTABLE NEW TLC | STAPLE, IMPLANTABLE | GDW | ETHICON ENDO-SURGERY, LLC. | NA | UNK |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |