FDA Adverse Event Malfunction Summary report: N

75MM SELECTABLE NEW TLC

MDR report key: 3940970 · Received July 17, 2014

Report

Report Number
3005075853-2014-05029
Event Type
Malfunction
Date Received
July 17, 2014
Date of Event
May 20, 2014
Report Date
May 27, 2014
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
GDW
PMA / PMN Number
K092577
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). FIRING MECHANISM. THE ANALYSIS RESULTS FOUND THAT THE NTLC75 DEVICE WAS RECEIVED WITH THE FIRING MECHANISM DAMAGED AS THE FIRING KNOB WAS DETACHED AND THE SLIP BLOCK ASSEMBLY WAS NOTED TO BE DAMAGED. THE DEVICE WAS RECEIVED WITH ONE RELOAD PRESENT. THE RELOAD WAS RETURNED WITH THE PROXIMAL 57 DRIVERS UP WITHOUT STAPLES, AND THE REMAINING DRIVERS DOWN WITH STAPLES PRESENT; THE SWING TAB IN THE LOCKED POSITION, WHICH INDICATES THAT THE DEVICE'S FIRING CYCLE WAS INTERRUPTED. THE DAMAGE TO THE FIRING KNOB AND SLIP BLOCK ASSEMBLY IS CONSISTENT WITH HIGH (OUTSIDE INDICATED USE) STAPLE FORMING FORCES; HOWEVER THERE IS INSUFFICIENT EVIDENCE TO DETERMINE THE CAUSE OF THE HIGHER LOADS. IT SHOULD BE NOTED THAT THE CARTRIDGE RELOAD IS DESIGNED TO LOCKOUT, AS A SAFETY FEATURE, IF ANY STAPLES HAVE BEEN FIRED FROM THE CARTRIDGE RELOAD. IF ENOUGH FORCE IS APPLIED THE DEVICE COULD BE DAMAGED. IN ADDITION FAILURE TO COMPLETE THE STROKE MAY RESULT IN INCOMPLETE STAPLE LINE. FOR ADDITIONAL INFORMATION PLEASE REFER TO THE INSTRUCTIONS FOR USE. DUE TO THE CONDITION OF THE DEVICE, NO FUNCTIONAL TEST COULD BE PERFORMED WITH IT. A BATCH RECORD REVIEW WAS PERFORMED AND THE BATCH HAD NO ANOMALIES NOTED DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN OPEN COLON PROCEDURE, ACCORDING TO THE SURGEON ON THE FOURTH FIRING THE STAPLER JAMMED. THE BLADE WOULD NOT MOVE FORWARD AND THEN THE HANDLE BROKE OFF. CASE COMPLETED WITH ANOTHER DEVICE OF THE SAME PRODUCT CODE. THERE WERE NO PATIENT CONSEQUENCES REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
419000 75MM SELECTABLE NEW TLC STAPLE, IMPLANTABLE GDW ETHICON ENDO-SURGERY, LLC. NA UNK

Patients

Seq Age Sex Outcome Treatment
1