PROMUS ELEMENT ?
Report
- Report Number
- 2134265-2014-04186
- Event Type
- Injury
- Date Received
- July 17, 2014
- Date of Event
- May 15, 2013
- Report Date
- June 23, 2014
- Manufacturer
- BOSTON SCIENTIFIC - GALWAY
- Product Code
- NIQ
- PMA / PMN Number
- SIMILAR
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR
- Reporter Occupation
- PHYSICIAN
Narratives
DEVICE IS COMBINATION PRODUCT. DEVICE EVALUATED BY MANUFACTURER: IT IS INDICATED THAT THE DEVICE WILL NOT BE RETURNED FOR EVALUATION; THEREFORE A FAILURE ANALYSIS OF THE COMPLAINT DEVICE COULD NOT BE COMPLETED. A REVIEW OF THE BATCH HISTORY, HISTORICAL TRENDING, AND SIMILAR COMPLAINT TRENDING REVIEW FOR THE PRODUCT FAMILY WILL BE CONDUCTED. IF THERE IS ANY FURTHER RELEVANT INFORMATION FROM THAT REVIEW, A SUPPLEMENTAL MEDWATCH WILL BE FILED. (B)(4).
(B)(4).
IT WAS REPORTED THAT DURING A PERCUTANEOUS TRANSLUMINAL ANGIOPLASTY TREATMENT PROCEDURE, IN-STENT RESTENOSIS OCCURRED. VASCULAR ACCESS WAS OBTAINED VIA THE FEMORAL ARTERY. THE 95% STENOSED, ECCENTRIC SHAPED, 20MM IN LENGTH, IN-STENT RESTENOSIS TARGET LESION WAS LOCATED IN THE MILDLY CALCIFIED AND MILDLY TORTUOUS, 2.5MM IN DIAMETER LEFT ANTERIOR DESCENDING ARTERY. THE LESION WAS PRE-DILATED. THIS 2.50X24MM PROMUS ELEMENT ¿ WAS DEPLOYED HOWEVER, ECHOCARDIOGRAM REVEALED IN-STENT RESTENOSIS. IT WAS NOTED THAT ANOTHER PTCA IS PENDING FOR TREATMENT OF THE RESTENOSIS. NO FURTHER PATIENT COMPLICATIONS WERE REPORTED AND THE PATIENTS STATUS IS STABLE.
IT WAS FURTHER REPORTED THAT PRE-DILATION WAS COMPLETED WITH A 2.5X15MM BALLOON CATHETER AND THIS 2.5X24MM PROMUS ELEMENT STENT WAS INFLATED TO 16 ATM. THE PROXIMAL SEGMENT OF THE STENT WAS POST DILATED WITH A 2.7X12MM BALLOON CATHETER AT 15 ATM.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 418382 | PROMUS ELEMENT ? | STENT, CORONARY, DRUG-ELUTING | NIQ | BOSTON SCIENTIFIC - GALWAY | H7493911324250 | 15307940 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 70 YR | Required Intervention |