TOTAL CARE BARIATRIC BED
Report
- Report Number
- 1824206-2014-01276
- Event Type
- Malfunction
- Date Received
- April 22, 2014
- Date of Event
- March 29, 2014
- Report Date
- March 29, 2014
- Manufacturer
- HILL-ROM, INC.
- Product Code
- FNL
- PMA / PMN Number
- K122473
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- OTHER
Narratives
THE HILL-ROM TECHNICIAN THOROUGHLY TESTED AND INSPECTED THE BED AND FOUND NO ISSUES. THE HEAD OF THE BED FUNCTIONED AS DESIGNED. PER THE HILL-ROM SERVICE MANUAL THE TOTAL CARE BED SYSTEM REQUIRES AN EFFECTIVE MAINTENANCE PROGRAM. WE RECOMMEND THAT YOU PERFORM A SEMI-ANNUAL PREVENTATIVE MAINTENANCE. PREVENTATIVE MAINTENANCE WILL MINIMIZE DOWNTIME DUE TO EXCESSIVE WEAR. TEST THE BED FUNCTION CONTROLS FOR PROPER OPERATION OF THE FUNCTION AND MOMENTARY OPERATION. TEST ALL LOCKOUT CONTROLS AND INDIVIDUALLY CHECK FOR PROPER OPERATION. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS DID NOT SHOW HILL-ROM PERFORMED ANY PREVENTATIVE MAINTENANCE ON THIS BED. IT IS UNK IF THE FACILITY PERFORMS PREVENTATIVE MAINTENANCE ON THEIR BEDS. THE HILL-ROM SERVICE TECHNICIAN PROVIDED AN INSERVICE ON THE BED OPERATIONS. BASED ON THIS INFO, NO FURTHER ACTION IS REQUIRED.
HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE HEAD SECTION OF THE BED MOVED UP WHEN IN THE THERAPY MODE. THE BED WAS LOCATED AT THE ACCOUNT. THERE WAS NO PT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 242650 | TOTAL CARE BARIATRIC BED | A/C POWERD ADJUSTABEL HOSPITAL BED | FNL | HILL-ROM, INC. | 1840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |