FDA Adverse Event Malfunction Summary report: N

TRANSTAR STRETCHER

MDR report key: 3940759 · Received April 22, 2014

Report

Report Number
3006697241-2014-00381
Event Type
Malfunction
Date Received
April 22, 2014
Date of Event
March 25, 2014
Report Date
March 25, 2014
Manufacturer
HILL-ROM DE MEXICO S DE RL DE CV
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN FOUND THE BRAKE PADS WERE NOT MAKING GOOD CONTACT WITH THE WHEEL. A SEARCH OF THE HILL-ROM MAINTENANCE RECORDS SHOWED HILL-ROM PERFORMED PREVENTATIVE MAINTENANCE ON THE THIS BED IN 2013. IT IS UNKNOWN IF THE FACILITY PERFORMED ANY OTHER PREVENTATIVE MAINTENANCE ON THIS BED. THE TECHNICIAN ADJUSTED THE BRAKE PADS TO RESOLVE THE ISSUE. BASED ON THIS INFORMATION, NO FURTHER ACTION IS REQUIRED.

Description of Event or Problem · 1

HILL-ROM RECEIVED A REPORT FROM THE ACCOUNT STATING THE BRAKES WERE NOT HOLDING. "THE BED WAS LOCATED IN THE BASEMENT DOWN BED AREA AT THE ACCOUNT." THERE WAS NO PATIENT/USER INJURY REPORTED. THIS REPORT WAS FILED IN OUR COMPLAINT HANDLING SYSTEM AS COMPLAINT #(B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
242657 TRANSTAR STRETCHER WHEELED STRETCHER FPO HILL-ROM DE MEXICO S DE RL DE CV 8000

Patients

Seq Age Sex Outcome Treatment
1