FDA Adverse Event Injury Summary report: N

840 VENTILATOR

MDR report key: 3940686 · Received June 25, 2014

Report

Report Number
8020893-2014-01490
Event Type
Injury
Date Received
June 25, 2014
Date of Event
May 26, 2014
Report Date
May 28, 2014
Manufacturer
COVIDIE, FORMERLY NELLCOR
Product Code
CBK
PMA / PMN Number
K970460
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NH, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE UNIT, AND VERIFIED THE REPORTED MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB), THE BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB'S), AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MFG SPECS. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER IN THE USA THAT, DURING PT USE, AN 840 VENTILATOR GRAPHIC USER INTERFACE (GUI) DISPLAY BECAME BLANK, AND IT CONTINUED CYCLING. THE PT WAS REMOVED FROM THE UNIT, AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO REPORTED PT INJURY.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
369560 840 VENTILATOR CBK COVIDIE, FORMERLY NELLCOR 840

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention