FDA Adverse Event
Injury
Summary report: N
840 VENTILATOR
MDR report key: 3940686
·
Received June 25, 2014
Report
- Report Number
- 8020893-2014-01490
- Event Type
- Injury
- Date Received
- June 25, 2014
- Date of Event
- May 26, 2014
- Report Date
- May 28, 2014
- Manufacturer
- COVIDIE, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NH, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) EVALUATED THE UNIT, AND VERIFIED THE REPORTED MALFUNCTION. THE CSE REPLACED THE GRAPHIC USER INTERFACE (GUI) CENTRAL PROCESSING UNIT (CPU) PRINTED CIRCUIT BOARD (PCB), THE BACKLIGHT INVERTER PRINTED CIRCUIT BOARDS (PCB'S), AND UPGRADED THE SOFTWARE TO THE CURRENT REVISION. THE UNIT PASSED ALL TESTS AND CALIBRATIONS, AND IT WAS OBSERVED TO BE OPERATING WITHIN MFG SPECS. (B)(4).
Description of Event or Problem · 1
IT WAS REPORTED BY A CUSTOMER IN THE USA THAT, DURING PT USE, AN 840 VENTILATOR GRAPHIC USER INTERFACE (GUI) DISPLAY BECAME BLANK, AND IT CONTINUED CYCLING. THE PT WAS REMOVED FROM THE UNIT, AND PLACED ON AN ALTERNATE VENTILATOR. THERE WAS NO REPORTED PT INJURY.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 369560 | 840 VENTILATOR | CBK | COVIDIE, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |