840 VENTILATOR
MDR report key: 3940661
·
Received June 25, 2014
Report
- Report Number
- 8020893-2014-01483
- Event Type
- Injury
- Date Received
- June 25, 2014
- Date of Event
- May 28, 2014
- Report Date
- May 28, 2014
- Manufacturer
- COVIDIEN, FORMERLY NELLCOR
- Product Code
- CBK
- PMA / PMN Number
- K970460
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION. THE CSE INSPECTED THE DEVICE AND REPLACED THE BREATH DELIVERY UNIT (BDU) PRINTED CIRCUIT BOARD (PCB). THE CSE UPGRADED THE SOFTWARE TO THE CURRENT REVISION AND PERFORMED THE EXTENDED SELF TEST (EST) SHORT SELF TEST (SST) AND THE PERFORMANCE VERIFICATION TEST (PVT); THE DEVICE OPERATED WITHIN THE MFG SPECS. (B)(4).
Description of Event or Problem · 1
A REPORT RECEIVED FROM (B)(6) STATED AN 840 VENTILATOR EXPERIENCED A DEVICE ALERT SUBSEQUENTLY CAUSING THE VENTILATOR TO STOP CYCLING WHILE BEING USED ON A PT. THE PT WAS REMOVED FROM THE DEVICE, MANUALLY VENTILATED THEN PLACED ON AN ALTERNATE VENTILATOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 369550 | 840 VENTILATOR | CBK | COVIDIEN, FORMERLY NELLCOR | 840 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |