FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3940431 · Received April 17, 2014

Report

Report Number
3008642652-2014-01167
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 10, 2014
Report Date
April 14, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF CHARGER/MODEM SN (B)(4) HAS BEEN COMPLETED. THE REPORT PROBLEM (DOES NOT LIGHT UP) WAS CONFIRMED. UPON EVALUATION, THERE WAS CONTAMINATION ON THE BATTERY BOARD. THE CAUSE OF THE INABILITY TO RECOGNIZE A BATTERY PACK IS THE CONTAMINATION. THE ROOT CAUSE OF THE CONTAMINATION COULD NOT BE POSITIVELY IDENTIFIED BUT WAS LIKELY LIQUID INGRESS OF AN UNK CONTAMINANT. NO ADVERSE EVENT RESULTED FROM THE CONTAMINATED CHARGER/MODEM. THE PT RECEIVED A REPLACEMENT CHARGER/MODEM.

Description of Event or Problem · 1

A PT SERVICE REPRESENTATIVE (PSR) CONTACTED ZOLL CUSTOMER SUPPORT WHILE ASSISTING A (B)(6) MALE PATIENT TO REPORT THAT THE PT'S CHARGE/MODULE WOULD NOT 'LIGHT UP'. THE PT WAS ISSUED A REPLACEMENT CHARGER/MODEM.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236051 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERGER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 72 YR