FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3940421
·
Received April 17, 2014
Report
- Report Number
- 3008642652-2014-01176
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- April 1, 2014
- Report Date
- April 14, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. UPON EVALUATION, THE CABLE CONNECTING THE REAR THERAPY ELECTRODE (TE) AND THE DISTRIBUTION NODE (DN)WAS PULLED FROM THE DN STRAIN RELIEF, DAMAGING INTERNAL WIRES. THE ROOT CAUSE OF THE DAMAGED CABLE AND WIRES CANNOT BE POSITIVELY IDENTIFIED BUT IS LIKELY EXCESSIVE FORCE PLACED ON THE CABLE. NO ADVERSE EVENT RESULTED FROM THE DAMAGE CABLE AND WIRES. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Description of Event or Problem · 1
REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, BELT SN (B)(4) FAILED INCOMING HI-POT TESTING. THE LAST PT TO USE THIS BELT DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 236058 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |