FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3940410
·
Received April 17, 2014
Report
- Report Number
- 3008642652-2014-01187
- Event Type
- Malfunction
- Date Received
- April 17, 2014
- Date of Event
- March 16, 2014
- Report Date
- April 15, 2014
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- DE, US
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (UNABLE TO BASELINE) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED AN ECG FALL-OFF TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO OUT-OF-TOLERANCE PINS 9 AND 16 ON PROCESSOR U700 ON THE DISTRIBUTION NODE PCA. THE ROOT CAUSE FOR THE DEFECTIVE U700 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE U700 COMPONENT. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.
Description of Event or Problem · 1
A ZOLL PATIENT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT SHE WAS UNABLE TO BASELINE A (B)(6) FEMALE PT. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 235613 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 73 YR |