FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3940410 · Received April 17, 2014

Report

Report Number
3008642652-2014-01187
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 16, 2014
Report Date
April 15, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
DE, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (UNABLE TO BASELINE) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT FAILED AN ECG FALL-OFF TEST. THE CAUSE FOR THE FAILURE WAS ISOLATED TO OUT-OF-TOLERANCE PINS 9 AND 16 ON PROCESSOR U700 ON THE DISTRIBUTION NODE PCA. THE ROOT CAUSE FOR THE DEFECTIVE U700 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE U700 COMPONENT. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A ZOLL PATIENT SERVICE REPRESENTATIVE (PSR) CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT SHE WAS UNABLE TO BASELINE A (B)(6) FEMALE PT. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
235613 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 73 YR