FDA Adverse Event Malfunction Summary report: N

LIFEVEST WCD 4000 SYSTEM

MDR report key: 3940408 · Received April 17, 2014

Report

Report Number
3008642652-2014-01186
Event Type
Malfunction
Date Received
April 17, 2014
Date of Event
March 15, 2014
Report Date
April 15, 2014
Manufacturer
ZOLL LIFECOR CORPORATION
Product Code
MVK
PMA / PMN Number
P010030
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DEVICE EVALUATION SUMMARY: DEVICE EVALUATION OF ELECTRODE BELT SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (SERVICE CODE 204) HAS BEEN CONFIRMED. UPON INVESTIGATION THE ELECTRODE BELT WAS UNABLE TO COMMUNICATE WITH A MONITOR. THE CAUSE FOR THE FAILURE WAS ISOLATED TO OPEN WIRES IN THE TRUNK CABLE. THE TRUNK CABLE SHOWED SIGNS OF PHYSICAL ABUSE. THE ROOT CAUSE FOR THE OPEN PULSE WIRE COULD OT BE POSITIVELY IDENTIFIED BUT WAS LIKELY EXCESSIVE FORCE. NO ADVERSE EVENT RESULTED FROM THE OPEN WIRES. THE PT RECEIVED A REPLACEMENT ELECTRODE BELT.

Description of Event or Problem · 1

A (B)(6) FEMALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT SHE WAS RECEIVING A SERVICE CODE 204. THE PT WAS ISSUED A REPLACEMENT ELECTRODE BELT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
236062 LIFEVEST WCD 4000 SYSTEM WEARABLE CARDIOVERTER DEFIBRILLATOR MVK ZOLL LIFECOR CORPORATION WCD 4000 NA

Patients

Seq Age Sex Outcome Treatment
1 68 YR